Storing a Microscope
ContentProper handling, maintenance and storage of the microscope are essential for proper functioning and life of the microscope.
The microscope should be placed and stored preferably in a box in a dry, dust-free and vibration-free environment, which is specially built in the laboratory (as shown in the figure below).

Figure: Storage of microscope in a wooden box; Source: Laboratory diagnosis by sputum smear microscopy
Important considerations for the storage of microscope
- When the microscope is not being used/ stored overnight, it should be covered or kept in a storage cupboard/box to keep it free from dust.
- Avoid exposing the microscope to direct sunlight.
- Avoid exposing the microscope to moisture as humidity may allow fungus to grow on the lens and cause rusting of the metal parts.
- The stored microscope must be kept warm with a light source to prevent growth of fungus. A bulb holder for a 15-watt bulb should be fixed on the rear wall of the storage cupboard, such that the microscope does not contact the bulb while storing or removing the microscope. The bulb should remain on when the microscope is stored inside the box
- An alternative may be to place plenty of dry blue silica gel into a shallow plate and place it in the box when the microscope is kept in it. Silica gel is blue in colour when it is dry, but when it becomes wet /absorbs moisture it turns pinkish. As soon as the silica gel becomes pink, it should be changed or heated until it turns blue again and then be reused.
Resources
Assessment
Question
Answer 1
Answer 2
Answer 3
Answer 4
Correct answer
Correct explanation
Page id
Part of Pre-test
Part of Post-test
Silica gel is blue in colour when it is dry.
True
False
1
Silica gel is blue in colour when it is dry.
Yes
Yes
Parts of a Microscope
ContentVideo fileWorking with a Microscope
ContentVideo fileCleaning a Microscope
ContentRegular cleaning of microscope is essential to ensure that it is optimally functioning. Cleaning is important to remove dirt, lens immersion oil and ensure disinfection. It involves cleaning the lenses, the body, the stage of the microscope and the light source. The microscope needs to be cleaned daily and the lenses need to be cleaned after each use.
Materials used for cleaning:
- Lint free cloth
- Lens paper
- Lens cleaning solution
- 70% ethanol
- Detergent solution
Procedure for Cleaning Different Parts of the Microscope
- The eyepiece shades, stage, focusing knob and nose piece are commonly touched during microscope operation, so these parts and the body of the microscope must be cleaned first to remove any stains with a neutral detergent /solution recommended by the manufacturer.
- To clean the microscope eyepiece, objective lens, surfaces of the condenser and the light exit glass, moisten a lens paper with 1 to 2 drops of lens cleaning solution and clean the lens/ glass with a circular/spiral motion (as shown in the figure below).
- Dry with a clean, dry piece of lens paper.
- Use the Spiral Wiping Technique: Wipe from the center to the periphery in a circular motion.
- The oil immersion lens should be cleaned after each use to ensure that immersion oil is not left on the surface.
- The microscope surface and parts (except the lenses) need to be disinfected using 70% ethanol.

Figure: Cleaning of Lenses using the Spiral Wiping Technique. Source: How to Clean the Microscope
Important Considerations for Cleaning Microscopes
- Do not blow air to remove the dirt.
- Do not wipe the lens with an ordinary cloth. All the lenses should be cleaned with dry lens paper/lint-free cloth.
- Never use spirit or alcohol or xylene to clean the lenses as these can damage them.
- Avoid using organic solvents that may damage plastic parts.
- Whenever possible, use the cleaning fluid recommended by the manufacturer.
Video fileResources
- Laboratory Diagnosis of Tuberculosis by Sputum Smear Microscopy, GLI, 2013.
- Module for Laboratory Technicians, CTD, 2005.
- How to Clean and Sterilise Your Microscope, Olympus, Life Science Solutions, 2020.
Assessment
Question
Answer 1
Answer 2
Answer 3
Answer 4
Correct answer
Correct explanation
Page id
Part of Pre-test
Part of Post-test
The lenses of the microscope should be cleaned with 70% alcoholic solution.
True
False
2
Lenses should never be cleaned with alcoholic solution and should always be cleaned with lens solution and lens paper.
Yes
Yes
At what frequency should the lenses of a microscope be cleaned? After each use Daily Weekly Monthly 1 The oil immersion lens should be cleaned after each use to ensure that the immersion oil is not left on the surface. Yes Yes Common issues with Microscope that require technical support
ContentRemovable parts in a microscope that need replacement include objectives, eyepieces, light bulbs, fuses. In case of repair/ technical support, only competent agency that handles maintenance of instruments should be contacted. Laboratory personnel should never attempt to dismantle any part of the microscope for repair.
Some of the common technical problems encountered, their likely causes and solutions are given in the table below.
Problems
Likely causes
Solutions
The viewing field is too dim
Condenser is too low
Condenser iris diaphragm is closedRaise condenser to correct its position
Open the diaphragm properlyDark shadows in the field that move as you turn the eyepiece
Surface of eyepiece is scratched
Eyepiece is dirtyReplace the eyepiece
Clean the eyepieceImage with the high-power objective is not clear
Slide is upside down
There is an air bubble in the oil
There is dirt on the objectiveTurn the slide over.
Move 100x lens quickly from side to side
Clean the lensImage with the low-power objective is not clear
There is oil on the lens
There is a layer of dust on the upper surface of the objectiveClean the lens
Mechanical stage is not moving, too stiff or does not stay up
Poor tension adjustment on the mechanical stage
Solidified lubricantsAdjust tension with adjustment knob
Needs serviceIn case, the viewing field is still dim and cloudy, consider the following possible causes and contact the competent personnel for cleaning/ repair:
- Massive growth of fungus on the lenses or prisms due to storage in a high-humidity environment
- Penetration of immersion oil between the lenses of the objective
- A damaged objective (as a result of careless focusing, dropping, rough changing of slides).
Video fileResources
1.
Module for Laboratory Technicians
2.
Laboratory diagnosis by sputum smear microscopy
Assessment
Question
Answer 1
Answer 2
Answer 3
Answer 4
Correct answer
Correct explanation
Page id
Part of Pre-test
Part of Post-test
What is the possible cause for dark shadows in the field?
The scratched eyepiece or dirty eyepiece
Broken lens
Solidified lubricants
Stage stuck
1
Dark shadows in field appear due to a scratched eyepiece or dirty eyepiece.
Yes
Yes
Annual Maintenance [AMC] of a Microscope
ContentThe binocular microscope requires considerable care in its use, regular cleaning, and protection from dust and fungal growth.
Annual maintenance of a microscope includes:
- Checking for broken or damaged parts and ensuring that the lenses, mirrors and other light-conducting surfaces are clean.
- Cleaning the lenses first for dirt with a blower brush, then wiping with lens solution on a lens paper/ 70% alcohol on the lens paper.
- Removing eyepieces or objectives from their fixation holes for cleaning.
- Properly cleaning the lower lens of the condenser after removal from its fixing.
- Cleaning the slide holder after removing from the mechanical stage.
- Cleaning for any deposition of immersion oil and fungal growth on the prisms in the binocular tube and eyepieces
- Changing electrical services including bulbs and fuses.
- Greasing/ lubrication of movable parts.
National Tuberculosis Elimination Programme (NTEP) has earmarked INR 2000 for Annual Maintenance [AMC] per binocular microscope including spare parts and repairs.
Resources
- Module for Laboratory Technicians.
- Laboratory diagnosis by sputum smear microscopy.
- Norms and Basis of Costing.
Assessment
Question
Answer 1
Answer 2
Answer 3
Answer 4
Correct answer
Correct explanation
Page id
Part of Pre-test
Part of Post-test
What is the AMC cost for each binocular microscope under NTEP?
INR 1000
INR 5000
INR 3000
INR 2000
4
NTEP has earmarked INR 2000 for Annual Maintenance (AMC) per binocular microscope, including spare parts and repairs.