Common Problems and Troubleshooting Related to Autoclaves in TB Lab
Table: Common Problems and Troubleshooting Related to Autoclaves in TB Lab
|
SR NO
|
PROBLEM
|
CAUSES
|
SOLUTION
|
|---|
|
1
|
Drain is clogged or exhaust taking more time than normal
|
There may be clogging of drain valve dust
|
- Clean the chamber
- Drain the filter
|
|
2
|
Scaling and salt deposition
|
- Cleaning is not proper
- There is moisture inside the instrument
- Exposure to chemicals like acids & detergents
|
- Cleaning must be improved and proper
- Check the sterilizer for efficient drying
- Store autoclave in a dry area
- Do not expose instruments to chemicals like acids and detergents. If exposure occurs, rinse thoroughly after contact
- Use only distilled water in the boiler
|
|
3
|
Liquid container caps blowing off
|
The sterilizer is exhausting too rapidly
|
Use slow exhaust cycle during autoclaving
|
|
4
|
When the door of the autoclave is open, the solution is still boiling
|
The door of an autoclave is opened too quickly
|
- Do not open an autoclave door until the temperature gauge is below 85⁰C and the pressure gauge is at "0"
- Do not touch or move a load, when the solution is boiling
|
|
5
|
Indicator of an autoclave shows sterilization not complete
|
- Load in an autoclave chamber may be too large, too dense, or improperly loaded
- The time provided for an autoclaving is not sufficient for the load
- Autoclave may be malfunctioning
|
- Don’t overload an autoclave with the material
- Load must be placed properly in an autoclave chamber
- Increase the time given for autoclaving
- Strips may need to be changed
|
|
6
|
Steam leakage
|
- The gasket may be worn out
- The door of an autoclave may be closed improperly
|
- The gasket needs to be changed
- The door needs to be closed properly
|
|
7
|
Chamber door won’t open
|
- There may be a vacuum in the chamber
- The door lock clutch of an autoclave may be jammed
- The gasket of an autoclave may be sticking to the door frame
|
- Follow manufacturer instructions properly during each cycle of autoclave
- Contact service engineer, if the problem persists
|